P9235 SPECTROANALYSER

The Cussons Spectroanalyser 3 has been developed to measure low concentrations of nitrogen as an impurity in certain rare gases, and has found particular applications with high purity argon.

APPLICATIONS
  • Measurement of trace amounts of nitrogen in high purity argon production and user systems
  • Quality control of high purity/ultra-high purity argon cylinders
FEATURES
  • 0-100 mV and 4-20 mA recorder outputs
  • High and low level alarm contacts
  • Analyser electronic self-check facility
  • Design eliminates background interference
 TECHNICAL SPECIFICATION
  • Ranges: 0-10 vpm and 0-100 vpm Nitrogen in Argon
  • Display: Analogue meter
  • Limit of detection: Less than 1 vpm
  • Sensitivity: Less than 1 vpm
  • Zero drift: Less than 0.1 vpm per ºC
  • Cycle/response time: Continuous/90% within 60 secs
  • Accuracy: Depending on calibration gas available as it is a comparative method that is employed
  • Sample flow rate: 2 litres per minute
  • Sample inlet temperature: 10ºC minimum, 30ºC max.
  • Sample inlet pressure: Optimum 0.35 to 1.05 bar g; maximum 3.5 bar g
  • Sample outlet pressure: Maximum 0.15 bar g
  • Sample pipework: 6mm or 1/4 OD stainless steel and PTFE
  • Analogue output: 4-20 mA and 0-100 mV
  • Alarm contacts: Adjustable high and low alarms with both normally open and closed voltage free contacts rated at 250V, 5A, AC non inductive
  • Electrical supply: 110/220V, 60 Hz or 220/240V, 50 Hz single phase

#Spectroanalyser #argon #calibrationgas #purityargon #nitrogen #Analyserelectronic

For more information:http://www.cussons.co.uk/